Traditional Calibration

Gross
Live Weight Reading

Step 1 -
Enter today's date (optional)
Calibration Date

Step 2 -
Select the load cell sensitivity.
Loadcell Sensitivity

Step 3 -
Enter the Calibration Motion Tolerance, used to ignore motion in the scale during the calibration process. Start with a value 3 times the graduation size, if calibration fails due to motion, consider increasing this value.
Cal Motion Tolerance

Step 4 -
Enter the Calibration Low Weight of the scale (this is also considered the dead load). If nothing is on the scale, this value will be zero.
Cal Low Weight
Step 5 -
Set the calibration parameters.
Step 6 -
Ensure that the weight (if any) on the scale is equal to the Cal Low Weight set in the previous step then click the Do Cal Low button.
Step 7 -
Enter the calibration span weight
Span Weight

Step 8 -
Ensure that the span weight on the scale is equal to the Span Weight set in the previous step then click the Do Cal High button.
Once you have successfully calibrated, you are done with this section. Return to the Main Wizard Page